1. Overview of 4-level steam indicator strips
The 4-level steam indicator strips are an important tool for sterilization process validation. They are made of 240 g/m2 cardboard and special steam sterilization indicators. These indicator strips are designed for high-pressure steam sterilizers and comply with ISO 11140-1 and EN 867 international standards. They can visually display whether the sterilization process has reached the preset critical conditions (temperature and time).
Product Features:
Autoclavable: Yes
Material: High-quality cardboard + special indicator
Color change: blue → black (after reaching sterilization conditions)
Standard sterilization parameters: 121°C/15 minutes or 134°C/3.5 minutes
2. Operation Guide
- Preparation before use
Check the integrity of the package to ensure that the indicator strip is not damaged
Confirm that the sterilized items are suitable for steam sterilization
Understand the type of sterilizer and preset parameters (temperature and time)
- Correct placement method
Place the indicator strip inside each sterilization package
For large sterilization packages, it is recommended to place it at the geometric center
Avoid direct contact with the inner wall of the sterilization package or metal instruments
Place multiple indicator strips at different positions in the sterilizer to monitor uniformity
- Sterilization process monitoring
Run the sterilization program
Observe whether the sterilizer parameters meet the preset conditions
Record the temperature and time data during the sterilization process
- Post-sterilization inspection
Immediately check the color change of the indicator strip after removing the sterilization package
Complete color change (blue → black) indicates that the sterilization conditions have been reached
Partial color change or no color change indicates that there may be problems in the sterilization process
Record the inspection results and save the records
3. Precautions
Storage conditions:
Temperature: 15-30°C
Humidity: <70%RH
Avoid direct sunlight
Keep away from chemical reagents
Usage restrictions:
Cannot be used for dry heat sterilization or chemical sterilization
Cannot replace biological indicators
Cannot be used as the only criterion for judging sterilization effect
Quality control:
Regularly check the perfor
Safety protection:
mance of the sterilizer
Use other monitoring methods (such as biological indicators) in conjunction
Establish a complete sterilization monitoring record system
Wear gloves when handling discolored indicator strips
Dispose of used indicator strips according to medical waste disposal procedures
4. FAQ
Q1: Why does the indicator strip not change color completely?
A: Possible reasons include:
Sterilization temperature does not reach the preset value
Insufficient sterilization time
Insufficient steam saturation
Improper placement of the indicator strip
Sterilizer failure
Q2: Can the indicator strip be reused after it changes color?
A: No. The Level 4 steam indicator strip is a one-time use product. After the color changes, its monitoring function is completed and it cannot be reused.
Q3: Which sterilizers can the indicator strip be used for?
A: Applicable to all high-pressure steam sterilizers that meet the standards, including:
Gravity displacement sterilizer
Pre-vacuum sterilizer
Pulsating vacuum sterilizer
Q4: How long is the shelf life of the indicator strip?
A: Usually 2-3 years (specific details are subject to the product packaging label), and it should be stored in a dry, light-proof, and temperature-appropriate environment.
Q5: Will the color of the indicator strip fade after it changes color?
A: Under normal conditions, the black color after the color change is stable and will not fade. However, it should be avoided from long-term exposure to strong light.